7. Post-recovery operations and calibrations#
Immediately following recovery, rinse the HydroC with freshwater and then let HydroC run in air for 1 hour to dry membrane and internal gas circuitry.
Return the HydroC to the manufacturer for a post-deployment calibration to allow span drift (changes in the sensor’s concentration dependent characteristics) correction during post processing using pre- and post-calibration coefficients interpolated over the deployment [Fietzek et al., 2014].